LC/MS Analysis of Phosphatidylethanol Metabolites on Ascentis® Express RP-Amide

LC/MS Analysis of Phosphatidylethanol Metabolites on Ascentis® Express RP-Amide application for LC-MS

Materials

related product

Product No.
Description
Pricing

Ascentis® Express RP-Amide, 2 μm UHPLC Column

2 μm particle size, L × I.D. 5 cm × 2.1 mm

CONDITIONS

column

Ascentis Express RP-Amide, 5 cm x 2.1 mm I.D., 2.0 μm particles (51569-U)

gradient

5 mM ammonium formate in [A] water; [B] methanol; [C] acetonitrile; (10:45:47, A:B:C)

flow rate

0.5 mL/min

pressure

2745 psi (189 bar)

column temp.

50 °C

detector

MS, ESI-, combined SIR: 699.7, 701.7 m/z

injection

2 μL

sample

500 ng/mL in 50:50, water:methanol

Description

Analysis Note

Phosphatidylethanol and its metabolites in blood are markers of ethanol consumption. Shown here is the rapid resolution of the metabolites on an Ascentis® Express RP-Amide column.