Skip to Content
Merck
CN

UHPLC/MS Analysis of Heroin and Related Compounds on Titan C18

UHPLC/MS Analysis of Heroin and Related Compounds on Titan C18 application for UHPLC

Materials

related product

Product No.
Description
Pricing

Diazepam solution

analytical standard, for drug analysis

Papaverine hydrochloride

powder

Noscapine hydrochloride hydrate

analytical standard

Codeine solution

1 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Alprazolam solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Flunitrazepam solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Heroin solution

1.0 mg/mL in acetonitrile, ampule of 1 mL, certified reference material, Cerilliant®

Morphine solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Titan C18 5 cm x 2.1 mm I.D., 1.9 μm particles (577122-U)

mobile phase

[A] 5 mM ammonium acetate in 95:5, water:acetonitrile; [B] 5 mM ammonium acetate in 5:95, water:acetonitrile

gradient

0 to 100% B in 3 min, held at 100% B for 1 min

flow rate

0.6 mL/min

pressure

6770 psi (467 bar)

column temp.

35 °C

detector

MS-TOF, XIC

injection

0.5 μL

sample

1 μg/mL in 95:5, water:methanol

Description

Analysis Note

This method shows heroin and related compounds separated on a Titan C18 column and MS detection. The highest quality LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.

Legal Information