产品名称
Millipore Express® SHC,Opticap® XL囊式, inlet connection diam. 3/4 in., Opticap® XL 600, pore size 0.5/0.2 μm, cartridge nominal length 6.5 in. (16.5 cm)
material
polyethersulfone
polyethylene support
polypropylene (gamma-stable)
polypropylene housing
polypropylene vent cap (Gamma stable)
polysulfone
silicone seal
reg. compliance
meets FDA Indirect Food Additive requirements cited in 21 CFR 177-182 (all component materials)
sterility
non-sterile
sterilization compatibility
gamma compatible
product line
Opticap® XL 600
feature
gamma compatible
hydrophilic
manufacturer/tradename
Millipore Express®
parameter
≤3.7 mL/min air diffusion at 2.8 bar (40 psig) and 23 °C (in water)
1.0 bar max. differential pressure (15 psid) at 80 °C (Forward)
1.0 bar max. inlet pressure (15 psi) at 80 °C
100 psi max. differential pressure (6.9 bar) (intermittent)
2.1 bar max. differential pressure (30 psid) at 25 °C (Reverse; intermittent)
2.75 bar max. inlet pressure (40 psi) at 60 °C
25 °C max. inlet temp.
5.5 bar max. differential pressure (80 psid) at 25 °C (Forward)
5.5 bar max. inlet pressure (80 psi) at 23 °C
6.9 bar max. differential pressure (100 psid) at 25 °C (Forward; intermittent)
6.9 bar max. inlet pressure (100 psi) at 23 °C (intermittent)
80 psi max. differential pressure (5.5 bar) (continuous)
80 psig max. inlet pressure
technique(s)
sterile filtration: suitable
L
16.5 cm (6.5 in.)
W
2.2 in.
cartridge nominal length
6.5 in. (16.5 cm)
diam.
5.6 cm (2.2 in.)
filtration area
590 cm2
inlet connection diam.
3/4 in.
inlet to outlet W
16.5 cm (6.5 in.)
outlet connection diam.
9/16 in.
impurities
≤0.25 EU/mL bacterial endotoxins (LAL test, Aqueous extraction)
<0.25 EU/mL USP bacterial endotoxins (LAL test, sample aqueous extraction)
matrix
Millipore Express® SHC
pore size
0.5/0.2 μm pore size
bubble point
≥4000 mbar (58 psig), air with water at 23 °C
fitting
1/4 in. drain/vent hose barb (with double O-ring seal)
inlet sanitary flange
outlet hose barb
(19 mm (3/4 in.) Sanitary Flange Inlet and 14 mm (9/16 in.) Hose Barb Outlet)
Quality Level
正在寻找类似产品? 访问 产品对比指南
Analysis Note
在灭菌和3.0 L的受控水冲洗后,样品在25 °C下显示出低于500 ppb TOC(根据USP<64>)和低于1.3 µS/cm(根据USP<64>)。
使用ASTM® F838方法,对样品进行最小保留的缺陷短波单孢菌挑战浓度为 1 × 10⁷ CFU/cm² 的定量保留。
Disclaimer
General description
Other Notes
- 生物截留:微生物
- 作用机制:过滤(体积排阻)
- 应用:生物处理
- 预期用途:减少或去除微生物/生物负荷
- 使用说明:请参见《OPTICAP® XL Capsules and OPTICAP® XLT胶囊用户指南》中的安装部分
- 储存说明:干燥储存
- 处置声明:根据国家、省市和当地适用法规处置。
Packaging
Preparation Note
耐受45kGy伽马射线辐照剂量;123°C×60 min高压锅蒸汽灭菌,3次;不可在线蒸汽灭菌
Legal Information
我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.
联系客户支持