Skip to Content
Merck
CN

Analysis of Pesticides in Black Tea by GC/MS/MS on SLB® Using Supelclean Ultra Cleanup

Analysis of Pesticides in Black Tea by GC/MS/MS on SLB® Using Supelclean™ Ultra Cleanup suitable for GC/MS

Materials

accessory

Product No.
Description
Pricing

Visiprep SPE Vacuum Manifold

DL (Disposable Liner), 12-port model

Supel QuE

Certified Vial Kit, Low Adsorption (LA) QsertVial

related product

Product No.
Description
Pricing

Sodium sulfate

ACS reagent, ≥99.0%, anhydrous, granular

Magnesium sulfate

anhydrous, ReagentPlus®, ≥99.5%

used together

Product No.
Description
Pricing

Supel QuE

Supelclean Ultra 2400

CONDITIONS

extraction tube

Supel Que non-buffered extraction tube (55294-U)

sample preparation

sample pretreatment: 1 gm sample in 50 mL centrifuge tube, add 10 mL water

sample preparation

Add QC check solution to each extract prior to GC/MS/MS analysis.

clean-up tube

Supelclean Ultra 2400, 3 mL (54281-U)

clean-up process

Add approx. 100 mg of annhydrous sodium sulfate to top of Ultra cartridge. Condition cartridge with 3 mL of acetonitrile. 0.100 mL (Load 0.750 mL of sample extract. Elute with 8 mL of acetontrile containing 0.5% formic acid. Collect eluent and concentrate at 40 °C to a final volume of )

column

SLB®-5ms, 30 m x 0.25 mm I.D., 0.250 μm (58471-U)

oven

105 °C (3 min), 10 °C/min to 130 °C, 4 °C/min to 200 °C, 8 °C/min. to 300 °C (10 min)

inj. temp.

programmed: 100°C (.05 min), 720 °C/min to 280 °C (45 min)

detector

MRM

MSD interface

300 °C

scan range

MRM

Description

General description

Extraction tube: 4 mg/NaSO4, 1 g NaCl in 12 mL tube

Analysis Note

Sample Prep: QuEChERS, Dispersive

Legal Information

SLB is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC
Supelclean is a trademark of Sigma-Aldrich Co. LLC