Skip to Content
Merck
CN

HPLC Analysis of Horny Goat Weed Extract Adulterated with Erectile Dysfunction Drugs on Ascentis® Express Phenyl-Hexyl

HPLC Analysis of Horny Goat Weed Extract Adulterated with Erectile Dysfunction Drugs on Ascentis® Express Phenyl-Hexyl application for HPLC

Materials

related product

Product No.
Description
Pricing

Ascentis® Express Phenyl-Hexyl (2.7 μm) HPLC Column

L × I.D. 10 cm × 3 mm, pk of 1 ea

Hyperoside

primary reference standard

Sildenafil solution

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

Vardenafil dihydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

sample preparation

Solvent Desorption

extraction

Weigh approximately 100 mg of herbal supplement and add 10 mL of 50:50, water:ethanol

sample preparation

Sonicate 20 min

sample preparation

Filter with 0.22 μm PVDF (Z227498)

column

Ascentis Express Phenyl-Hexyl, 10 cm x 3.0 mm I.D., 2.7 μm particles (53345-U)

mobile phase

[A] water with 0.1% formic acid; [B] acetonitrile with 0.1% formic acid;

gradient

5% B held for 0.5 min; to 100% B in 9.5 min; 100% B held for 2 min

flow rate

0.4 mL/min

pressure

2030 psi (140 bar)

column temp.

35 °C

detector

UV, 250 nm

injection

2 μL

sample

Extract 10mg/mL before filtration, adulteration 20μg/mL in 50:50, water:ethanol

Description