HPLC Analysis of Metalaxyl Enantiomers on Astec® Cellulose DMP

HPLC Analysis of Metalaxyl Enantiomers on Astec® Cellulose DMP application for HPLC

Materials

related product

Product No.
Description
Pricing

Metalaxyl-M

PESTANAL®, analytical standard

Metalaxyl

PESTANAL®, analytical standard

Astec® Cellulose DMP Chiral HPLC Column

5 μm particle size, L × I.D. 25 cm × 4.6 mm

used together

Product No.
Description
Pricing

Ethyl alcohol, Pure

200 proof, anhydrous, ≥99.5%

CONDITIONS

column

Astec Cellulose DMP, 25 cm x 4.6 mm I.D., 5 μm particles (51099AST)

mobile phase

[A] ethanol; [B] hexane, (25:75, A:B)

flow rate

1.0 mL/min

pressure

740 psi (51 bar)

column temp.

25 °C

detector

UV, 220 nm

injection

10 μL

sample

1 mg/mL in ethanol

Description

Analysis Note

Metalaxyl is a 1:1 mixture of (R)- and (S)-2-[(2,6-dimethylphenyl)methoxyacetylamino]propionic acid methyl ester. Both enantiomers are fungicides used in agriculture, horticulture, and forestry. They act by inhibiting mycelial growth and spore formation. Metalaxyl-M is the biologically active (R)-enantiomer. Technical-grade metalaxyl-M consists of a minimum of 97% of the (R)-enantiomer and 3% of the (S)-enantiomer. Chiral LC on Astec Cellulose DMP was used to separate the two enantiomers.