跳转至内容
Merck
CN

G003054

HPLC Analysis of Phenolic Antioxidants on Ascentis® C8

application for HPLC

登录查看公司和协议定价

选择尺寸


技术服务
需要帮助?我们经验丰富的科学家团队随时乐意为您服务。
让我们为您提供帮助
技术服务
需要帮助?我们经验丰富的科学家团队随时乐意为您服务。
让我们为您提供帮助

技术

HPLC: suitable

试验参数

column: Ascentis C8, 15 cm x 4.6 mm I.D., 5 μm particles (581424-U)
mobile phase: [A] 5% acetic acid; [B] (50:50) methanol:acetonitrile
gradient: 30 to 100% B in 10 min; held for 5 min
flow rate: 2.0 mL/min
column temp.: 30 °C
detector: UV, 280 nm
injection: 10 μL
sample: as indicated in 50:50 acetonitrile:2-propanol

适用性

application for HPLC

应用

cleaning products
cosmetics
personal care
vitamins, nutraceuticals, and natural products

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany

分析物

说明

    Propyl gallate
    Ethoxyquin
    2′,4′,5′-Trihydroxybutyrophenone
    tert-Butylhydroquinone 10 μg/mL
    Nordihydroguaiaretic acid
    3-tert-Butyl-4-hydroxyanisole 10 μg/mL
    2,6-Di-tert-butyl-4-hydroxymethylphenol 10 μg/mL
    Octyl gallate 10 μg/mL
    Lauryl gallate 10 μg/mL
    3,5-Di-tert-butyl-4-hydroxytoluene, analytical standard
    查看所有结果 (10)