HPLC Analysis of Statins in Rat Plasma on Ascentis® Express C18 after SPE using Supel-Select HLB

application for SPE, application for LC-MS

登录查看公司和协议定价

技术

LC/MS: suitable
solid phase extraction (SPE): suitable

试验参数

sample preparation: SPE (Solid Phase Extraction)
sample/matrix: rat plasma spiked with provastin and atorvastin at 5 and 100 ng/mL
SPE tube/cartridge: Supel-Select HLB SPE, 30 mg/1 mL (54181-U)
condition: 0.5 mL methanol:acetonitrile (1:1, v/v); 0.5 mL DI water
sample addition: 0.5 mL
washing: 0.5 mL 5% methanol
elution: 0.5 mL methanol:acetonitrile (1:1, v/v)
column: Ascentis Express C18, 5 cm x 2.1 mm, 2.7 μm particles (53822-U)
eluate post-treatment: evaporate to dryness with nitrogen at 37 °C, reconstitute in 0.5 mL DI water
mobile phase: 0.1% acetic acid diluted in 60% methanol
flow rate: 0.2 mL/min
column temp.: 35 °C
detector: ABI 3200 QT; ESI(+), MRM (423.3/321.3, 423.3/101.0, 557.3/397.2 and 557.3/453.4)
injection: 5 μL

适用性

application for LC-MS
application for SPE

应用

clinical

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany
Supel is a trademark of Sigma-Aldrich Co. LLC

分析证书(COA)

输入产品批号来搜索 分析证书(COA) 。批号可以在产品标签上"批“ (Lot或Batch)字后找到。

已有该产品?

为方便起见,与您过往购买产品相关的文件已保存在文档库中。

访问文档库

难以找到您所需的产品或批次号码?

在网站页面上,产品编号会附带包装尺寸/数量一起显示(例如:T1503-25G)。请确保 在“产品编号”字段中仅输入产品编号 (示例: T1503).

示例

T1503
货号
-
25G
包装规格/数量

其它示例:

705578-5MG-PW

PL860-CGA/SHF-1EA

MMYOMAG-74K-13

1000309185

输入内容 1.000309185)

遇到问题?欢迎随时联系我们技术服务 寻求帮助

批号可以在产品标签上"批“ (Lot或Batch)字后面找到。

Aldrich 产品

  • 如果您查询到的批号为 TO09019TO 等,请输入去除前两位字母的批号:09019TO。

  • 如果您查询到的批号含有填充代码(例如05427ES-021),请输入去除填充代码-021的批号:05427ES。

  • 如果您查询到的批号含有填充代码(例如 STBB0728K9),请输入去除填充代码K9的批号:STBB0728。

未找到您寻找的产品?

部分情况下,可能未在线提供COA。如果搜索不到COA,可在线索取。

索取COA

我们的科学家团队拥有各种研究领域经验,包括生命科学、材料科学、化学合成、色谱、分析及许多其他领域.

联系技术服务部门