技术
gas chromatography (GC): suitable
solid phase extraction (SPE): suitable
试验参数
sample preparation: SPE (Solid Phase Extraction)
sample/matrix: cod liver oil spiked with PCBs (at 10 ng/g)
SPE tube/cartridge: Supelclean™ EZ-POP NP, 12 mL, Pk 20 (54341-U)
condition: 10 mL acetone, 10 min dry at 20" mercury
sample addition: 0.5 g sample, weighed directly on cartridge
elution: 15 mL acetonitrile
eluate post-treatment: back extract acetonitrile eluent with 2 x 10 mL portions of hexane, evaporate hexane extract to approximately 1 mL at 40 ºC under nitrogen
SPE tube/cartridge: Supelclean™ LC-Si, 100 mg/1 mL (504041)
condition: 4 mL acetone, 10 min dry at 20" mercury, then 4 mL hexane
sample addition: 1 mL of resulting extract from above
elution: 5 mL hexane
eluate post-treatment: evaporate to 1 mL at 40 ºC under nitrogen
column: SLB-5ms 20m x 0.18mm, 0.18 μm (28564-U)
oven: 75 °C (1 min), 12 °C/min to 340 °C (20 min)
inj. temp.: 250 °C
detector: μECD, 340 °C
carrier gas: hydrogen, 1.2 mL/min
injection: 1 μL, splitless (0.75 min)
liner: 4 mm I.D., split/splitless type, wool packed single taper FocusLiner™ design