  • Rapid sucrose monitoring in green coffee samples using multienzymatic biosensor.

Rapid sucrose monitoring in green coffee samples using multienzymatic biosensor.

Food chemistry (2018-03-20)
Miroslav Stredansky, Luca Redivo, Peter Magdolen, Adam Stredansky, Luciano Navarini
摘要

Amperometric biosensor utilizing FAD-dependent glucose dehydrogenase (FAD-GDH) for a specific sucrose monitoring in green coffee is described. FAD-GDH was co-immobilized with invertase and mutarotase on a thin-layer gold planar electrode using chitosan. The biosensor showed a wide linearity (from 10 to 1200 μM), low detection limit (8.4 μM), fast response time (50 s), and appeared to be O2 independent. In addition the biosensors exhibited a good operational (3 days) and storage (1 year) stability. Finally, the results achieved from the biosensor measurements of sucrose in 17 samples of green coffee (Coffea arabica, C. canephora and C. liberica) were compared with those obtained by the standard HPLC method. The good correlation among results of real samples, satisfactory analytical performance and simple use of the presented biosensor make it suitable for application in coffee industry.

材料
产品编号
品牌
产品描述

Sigma-Aldrich
吩嗪硫酸甲酯