Oxalic acid, oxamic acid and oxamide are potential impurities in some active pharmaceutical ingredients (API). The retention and separation of oxalic and oxamic acids are particularly challenging using conventional reversed-phase HPLC due to their high polarity. An ion-exclusion chromatography (IEC) method has been shown to provide good separation and sensitivity for the three oxalate-related impurities in a hydrophobic API matrix. The method uses a Dionex IonPac ICE-ASI column with 95/5 (v/v) 0.1% sulfuric acid/acetonitrile as the mobile phase and UV detection at 205 nm. Development and validation of this method are described.