  • Organs-on-Chips with combined multi-electrode array and transepithelial electrical resistance measurement capabilities.

Organs-on-Chips with combined multi-electrode array and transepithelial electrical resistance measurement capabilities.

Lab on a chip (2017-06-14)
Ben M Maoz, Anna Herland, Olivier Y F Henry, William D Leineweber, Moran Yadid, John Doyle, Robert Mannix, Ville J Kujala, Edward A FitzGerald, Kevin Kit Parker, Donald E Ingber
摘要

Here we demonstrate that microfluidic cell culture devices, known as Organs-on-a-Chips can be fabricated with multifunctional, real-time, sensing capabilities by integrating both multi-electrode arrays (MEAs) and electrodes for transepithelial electrical resistance (TEER) measurements into the chips during their fabrication. To prove proof-of-concept, simultaneous measurements of cellular electrical activity and tissue barrier function were carried out in a dual channel, endothelialized, heart-on-a-chip device containing human cardiomyocytes and a channel-separating porous membrane covered with a primary human endothelial cell monolayer. These studies confirmed that the TEER-MEA chip can be used to simultaneously detect dynamic alterations of vascular permeability and cardiac function in the same chip when challenged with the inflammatory stimulus tumor necrosis factor alpha (TNF-α) or the cardiac targeting drug isoproterenol. Thus, this Organ Chip with integrated sensing capability may prove useful for real-time assessment of biological functions, as well as response to therapeutics.

材料
产品编号
品牌
产品描述

Sigma-Aldrich
Triton X-100, laboratory grade
Sigma-Aldrich
聚二甲基硅氧烷, viscosity 1.0 cSt (25 °C)