GC Analysis of a QC Test Mix Before/After Repeat 60% Water Injections on Watercol 1460

GC Analysis of a QC Test Mix Before/After Repeat 60% Water Injections on Watercol™ 1460 suitable for GC

CONDITIONS

column

Watercol 1460, 30 m x 0.25 mm I.D., 0.20 μm (29695-U)

oven

110 °C

inj. temp.

250 °C

detector

TCD, 200 °C

carrier gas

helium, 26 cm/sec

injection

1 μL, 100:1 split

liner

4 mm I.D., split type, cup design

sample

6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL)

说明

分析说明

The "Watercol" series of ionic liquid capillary GC columns are characterized by their ability to produce a sharp peak shape for water, sharp enough so that water can be integrated and quantified. Sharp peak shapes are also produced for other small polar analytes. This application shows the ruggedness of Watercol 1460 to repeat injections of 60% water.

其他说明

Conditions for each 60% water run = oven: 60 °C; inj. temp.: 250 °C; detector: TCD, 200 °C; carrier gas: helium, 28 cm/sec; injection: 1 μL, 100:1 split; liner: 4 mm I.D., split type, cup design; sample: water at 60% (v/v) in ethanol

法律信息

Watercol is a trademark of Sigma-Aldrich Co. LLC