HPLC-FLD Analysis of Bisphenol A in Olive Oil using a 4.6 mm ID Chromolith® Column and Ethanol in Mobile Phase - Green HPLC

HPLC-FLD Analysis of Bisphenol A in Olive Oil using a 4.6 mm ID Chromolith® Column and Ethanol in Mobile Phase - Green HPLC

材料

used together

产品编号
说明
价格

酒精

gradient grade for liquid chromatography LiChrosolv®

for chromatography (LC-MS Grade) LiChrosolv®

Milli-Q® IQ 7000 超纯水净化系统

output: type 1 water (18.2 MΩ·cm), the most advanced Milli-Q® ultrapure (Type 1) water purification system that is intelligent, intuitive, and green.

相关产品

产品编号
说明
价格

双酚 A

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

Chromolith® HighResolution RP-18 封端

L × I.D. 100 mm x 4.6 mm HPLC column

Millex亲水PTFE针式过滤器

pore size 0.45 μm, diam. 13 mm, non-sterile

CONDITIONS

column

Chromolith® HighResolution RP-18 endcapped 100 x 4.6 mm (1.52022)

mobile phase

Water/Ethanol 51/49 (v/v)

gradient

isocratic

flow rate

1 mL/min

pressure

69 bar (1001 psi)

column temp.

30 °C

detector

FLD; Ex: 275 nm; Em: 305 nm

sample/matrix

Olive Oil spiked at 25 µg/g with BPA - Weigh 2.0 g of olive oil in a 15 mL screw-cap test tube, add 1 mL of the spike solution (50 µg/mL in, seal with screw cap and shake for 30 seconds. Add 4 mL of diluent and shake for 1 minute. Sonicate the solution for 15 minutes at room temperature and after that centrifuge it at 5000 rev/min for 25 min. Take an aliquot of the supernatant and put it through a 0.45 µm PTFE Filter.

injection

5 µL (A037)

说明

一般描述

Bisphenol A (BPA) is a precursor to important plastics and can be found in many different consumer goods, e.g. plastic bottles. It is soluble in organic solvents and slightly soluble in water. Due to this characteristic leaching of BPA from plastic bottles into the bottle content can be observed. BPA is a xenoestrogen with the ability to mimic the human hormone estrogen.

分析说明

This application work is focused on combining HPLC with more sustainable solvents. Therefore, the analysis of BPA content in olive oil was developed by using ethanol as an organic modifier in the mobile phase and as extraction solvent.
For the HPLC separation, a matrix tolerant Chromolith® HighResolution RP-18 endcapped column 100 x 4.6 mm I.D. combined with fluorescence detection (FLD) was applied. A spiked olive oil sample (25 µg/g) was only extracted ethanol and filtered through a 0.45 µm PTFE Filter.

其他说明

App_727