GC Analysis of a QC Test Mix Containing Water on Watercol 1910, 96 °C

GC Analysis of a QC Test Mix Containing Water on Watercol 1910, 96 °C suitable for GC

材料

相关产品

产品编号
说明
价格

Watercol 1910 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

Watercol 1910, 30 m x 0.25 mm I.D., 0.20 μm (29711-U)

oven

96 °C

inj. temp.

250 °C

detector

TCD, 200 °C

carrier gas

helium, 26 cm/sec

injection

1 μL, 100:1 split

liner

4 mm I.D., split type, cup design

sample

6 analytes in acetone, each at 1 mg/mL except water (at 0.5 mg/mL)

说明

分析说明

The "Watercol" series of ionic liquid capillary GC columns are characterized by their ability to produce a sharp peak shape for water, sharp enough so that water can be integrated and quantified. Sharp peak shapes are also produced for other small polar analytes. This application shows a QC test mix containing water on Watercol 1910 at 96 °C isothermal.