Merck
CN

GC Analysis of Class 1 Residual Solvents on Equity-1 after SPME using 100 μm PDMS Fiber

GC Analysis of Class 1 Residual Solvents on Equity™-1 after SPME using 100 μm PDMS Fiber suitable for GC

材料

SPME 光纤

产品编号
说明
价格

SPME 纤维组件聚二甲基硅氧烷 (PDMS)

df 100 μm(PDMS, needle size 24 ga, for use with manual holder

分析色谱柱

产品编号
说明
价格

Equity-1 毛细管GC色谱柱

L × I.D. 10 m × 0.20 mm, df 1.20 μm

CONDITIONS

sample/matrix

2 mL water + sodium chloride at 25% containing each analyte at 1 ppm, in a 4 mL vial

SPME fiber

100 μm PDMS, 24 ga

extraction

heated headspace, 50 °C for 5 min

desorption process

250 °C for 3 min

column

Equity-1, 10 m x 0.20 mm I.D., 1.2 μm (28043-U)

oven

40 °C (0.75 min), 20 °C/min to 200 °C (10 min)

inj. temp.

250 °C

detector

FID

carrier gas

helium, 35 cm/sec @ 40 °C

injection

5:1 split

liner

0.75 mm I.D. SPME type

说明

法律信息

Equity is a trademark of Sigma-Aldrich Co. LLC