跳转至内容
Merck
CN

GC Analysis of Class 2 Residual Solvents on VOCOL® after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Class 2 Residual Solvents on VOCOL® after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

材料

used together

产品编号
说明
价格

聚丙烯酸酯 (PA) 固相微萃取萃取头

85 μm Polyacrylate, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, white hub

CONDITIONS

sample/matrix

2 mL water + sodium chloride at 25% (pH 11) containing each analyte at 5 ppm, in a 4 mL vial

SPME fiber

85 μm Polyacrylate (57304)

extraction

heated headspace, 60 °C for 5 min

desorption process

250 °C for 3 min

column

VOCOL, 10 m x 0.20 mm I.D., 1.2 μm (24129-U)

oven

40 °C (0.75 min), 20 °C/min to 200 °C (10 min)

inj. temp.

250 °C

detector

FID

carrier gas

helium, 35 cm/sec @ 40 °C

injection

5:1 split

liner

0.75 mm I.D. SPME type

说明

法律信息

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany