Merck
CN

GC Analysis of Phenols in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber

GC Analysis of Phenols in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

材料

相关产品

产品编号
说明
价格

聚丙烯酸酯 (PA) 固相微萃取萃取头

df 85 μm(PA, for use with manual holder, needle size 24 ga

CONDITIONS

sample/matrix

50 ppb phenols in water, 27% NaCl, pH 2

SPME fiber

polyacrylate, 85 μm (57304)

extraction

20 min immersed in water with rapid stirring

desorption process

splitless, 280 °C (closed 3 min)

column

5% Phenyl Polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)

oven

40 °C (4 min), 12 °C/min to 260 °C

detector

MS, Scan Range m/z = 45-465 at 0.6 sec/scan

carrier gas

helium, 40 cm/sec at 40 °C