SPE (Solid Phase Extraction)
acidified drinking water
ENVI-18 DSK SPE Disk, 47 mm (57171)
5 mL dichloromethane; 5 mL methanol; 5 mL DI water
5 mL water sample
2 x 5 mL acetonitrile rinsing internal chamber of sample reservoir thoroughly
5% phenyl polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
40 °C to 160 °C (3 min), then to 300 °C (3 min) at 6 °C/min
MS, scan range m/z = 45-450
helium, 33 cm/sec
1 μL split/splitless, 1 min delay
Sample Pre-treatment:
Adjust 1 L drinking water to pH <2 with 6 N HCl. Add 5 mL methanol and mix thoroughly
ENVI is a trademark of Sigma-Aldrich Co. LLC