跳转至内容
Merck
CN

GC Analysis of Volatiles in Water on VOCOL after SPME using 100 μm PDMS Fiber

GC Analysis of Volatiles in Water on VOCOL after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

CONDITIONS

sample/matrix

50ppb each analyte in 1.8 mL saturated salt water

SPME fiber

polydimethylsiloxane, 100 μm film (57300-U)

extraction

5 min immersed in water with rapid stirring

desorption process

230 °C, splitless (closed 3 min)

column

VOCOL, 10 m × 0.20 mm I.D., 1.2 μm (24129-U)

oven

40 °C (0.75 min), 20 °C/min to 160 °C

detector

FID, 260 °C

carrier gas

helium, 40 cm/sec