GC Analysis of Water in Gasoline (Unspiked) on Watercol 1910

GC Analysis of Water in Gasoline (Unspiked) on Watercol™ 1910 suitable for GC

材料

相关产品

产品编号
说明
价格

Watercol 1910 Capillary GC Column

L × I.D. 30 m × 0.25 mm, df 0.20 μm

CONDITIONS

column

Watercol 1910, 30 m x 0.25 mm I.D., 0.20 μm (29711-U)

oven

45 °C (4 min), 5 °C/min to 75 °C, 10 °C/min to 190 °C

inj. temp.

250 °C

detector

FID, 250 °C; TCD, 200 °C

carrier gas

helium, 20 cm/sec (measured in each column at an oven temperature of 135 °C)

injection

1 μL, 50:1 split

liner

4 mm I.D., split/splitless type, wool packed single taper FocusLiner design

sample

(unleaded gasoline)

说明

分析说明

Residual moisture in fuel is very undesirable. This is because it reduces the heat of combustion which increases engine stress, corrodes fuel system components, allows gelling of fuel in cold temperatures through ice crystal formation and nucleation, and accelerates the growth of microbial colonies which can clog fuel systems. The "Watercol" series of ionic liquid capillary GC columns are characterized by their ability to produce a sharp peak shape for water, sharp enough so that water can be integrated and quantified.

This application shows the suitability of Watercol 1910 for the analysis of water in gasoline. In addition to identification of water, the column′s selectivity allowed for the identification of the BTEX analytes. The water and ethanol peaks were confirmed by injecting a series of standards (water in ethanol). The BTEX peaks were confirmed by analysis of neat materials. The unique combination of stationary phase:analyte interactions also allowed the elution of the heavy components (e.g. methylnaphthalenes) in a reasonable time (<19 minutes).

其他说明

Two identical columns were installed in separate inlets, each going to a separate detector.

法律信息

FocusLiner is a trademark of SGE Analytical Science Pty Ltd.
Watercol is a trademark of Sigma-Aldrich Co. LLC