US EPA Method 325: GC Analysis of Volatile Organic Compounds on SPB®-HAP after Thermal Desorption using FLM Carbopack X Tube

US EPA Method 325: GC Analysis of Volatile Organic Compounds on SPB®-HAP after Thermal Desorption using FLM Carbopack™ X Tube suitable for GC, suitable for air monitoring

材料

配件

产品编号
说明
价格

ATIS 吸附管进样器系统 110V

用于 Luer 针头和带 TLL 锁定器进样针的 Hamilton® 针头

Metal Hub (N), needle size 22 ga (cone tip, side-port), needle L 51 mm (2 in.), pkg of 6 ea

Hamilton® Samplelock 注射器

1010SL, volume 10 mL, needle size 22 ga (bevel tip), needle L 51 mm (2 in.)

Luer 公接头/可拆卸针头的接头,件

converts large hub RN termination (250 μL or larger) to a Male luer connection.

CONDITIONS

sample/matrix

14 Day Storage Time after Spiking with Gas Mix

thermal desorption tube

FLM Carbopack X 1/4" (6.35 mm O.D. X 3.5" (89mm) Long Stainless Steel TD Tube (28686-U)

dry purge

50 mL/min @ Ambient for 1 min

desorption preheat

FLM Carbopack-X Tube Desorption: 50 mL/min @ 330°C for 5 min

desorption process

Refocusing Trap: 7 mL/min @ 330°C for 5 min

bake

n/a

transfer line/valve temp.

175 °C

column

SPB-HAP. 60 m x 0.32 mm I.D., 4.0μm (25020-U)

oven

40° C (5 min), 5°C/min to 100° C, 15°C/min to 220° C (12 min)

detector

MSD

MSD interface

230°C

scan range

Full Scan. m/z 35-265

carrier gas

helium, 2 mL/min @ambient

injection

Thermal Desorption

liner

N/A

说明

分析说明

Carbopack-X TD spiked with 12.7 mL of the gas mix using the ATIS Adsorbent Tube Injector System. Swagelok® endcaps were placed on the tube. The spiked tube was stored at ambeint temperature for 14 days prior to analysis.