US EPA Method 524.2: GC Analysis of Volatiles on SPB-624 after Purge & Trap using "K" Trap

US EPA Method 524.2: GC Analysis of Volatiles on SPB-624 after Purge & Trap using "K" Trap suitable for GC

材料

分析色谱柱

产品编号
说明
价格

SPB®-624毛细管GC色谱柱

L × I.D. 30 m × 0.25 mm, df 1.40 μm

标准

产品编号
说明
价格

EPA 524.2 Rev.4 更新混标

2000 μg/mL in methanol, analytical standard

EPA 524 内标混标

certified reference material, 2000 μg/mL each component in methanol

EPA 524.2 Rev.4 更新酮类混标

certified reference material, 2000 μg/mL in methanol: water (9:1)

EPA 502/524 挥发性有机物校准混标 A(无气体)

certified reference material, 200 μg/mL each component in methanol, ampule of 1 mL

挥发性有机化合物混标 6

certified reference material, 2000 μg/mL each component in methanol

相关产品

产品编号
说明
价格

吹扫/捕集 K Vocarb® 3000

for use with OI Analytical Eclipse 4660, 4560

CONDITIONS

sample/matrix

each analyte at 10 ppb in 25 mL water

purge trap

VOCARB 3000 "K" (24940-U)

dry purge

2 min

desorption preheat

205 °C

desorption process

40 mL/min at 210 °C for 2 min

bake

260 °C for 10 min

transfer line/valve temp.

110 °C

sample preparation

purge is 40 mL/min at 25 °C for 11 min

column

SPB-624, 30 m x 0.25 mm I.D., 1.4 μm (24255)

oven

40 °C (2 min), 7 °C/min to 135 °C, 30 °C/min to 230 °C (3 min)

carrier gas

helium, 1.2 mL/min

injection

30:1 split

liner

0.75 mm I.D. SPME

说明

分析说明

Method 524.2 describes the analysis of volatile organic compounds in surface, ground, and drinking water. This application illustrates the analysis of the compounds listed in this method on a 30m x 0.25 mm I.D., 1.4 μm SPB-624 column using purge and trap / GC-MS.