US EPA Method 608/8081/OLM04.2 PEST: GC Analysis of Organochlorine Pesticides on SLB®-5ms, 340 °C Final Temp.

US EPA Method 608/8081/OLM04.2 PEST: GC Analysis of Organochlorine Pesticides on SLB®-5ms, 340 °C Final Temp. suitable for GC

材料

相关产品

产品编号
说明
价格

SLB ®-5 ms 毛细管 GC 色谱柱

L × I.D. 30 m × 0.25 mm, df 0.25 μm

EPA 8081 农药混标

certified reference material, 200 μg/mL in hexane: toluene (1:1), ampule of 1 mL

CONDITIONS

column

SLB-5ms, 30 m x 0.25 mm I.D., 0.25 μm (28471-U)

oven

100 °C (2 min), 15 °C/min to 340 °C (5 min)

inj. temp.

250 °C

carrier gas

helium, 0.9 mL/min

injection

1μL, splitless (0.75 min)

liner

4 mm I.D., split/splitless type, wool packed single taper FocusLiner design

sample

200 ppm organochlorine pesticide standard mix (CRM46845) containing 22 analytes, diluted to 50 ppb with hexane

detector

μECD, 340 °C

说明

分析说明

The analysis of low levels of 20 organochlorine pesticides and 2 surrogate compounds by gas chromatography-electron capture detector (GC-ECD) is routinely performed by analysts in environmental laboratories. Methods require the analysis of each sample extract on two columns with differing selectivity for confirmation of identifications. The SLB-5ms/SLB-35ms pair is an effective column set because all 22 analytes are resolved between both columns using the same run conditions.

This applicatin shows the separation achieved on SLB-5ms. Peaks are well integrated and baselines are exceptionally stable considering the high final oven temperature (340 °C). Minimal baseline rise is detected, and the last analyte, decachlorobiphenyl, is eluted in less than 20 minutes.