US EPA Method 8082: GC Analysis of Aroclor 1254 (PCB) on SLB®-5ms and Equity-1701 (15 m x 0.10 mm I.D., 0.10 μm), Fast GC Analysis

CONDITIONS
column
SLB-5ms, 15 m x 0.10 mm I.D., 0.10 μm (28466-U)
column
Equity-1701, 15 m x 0.10 mm I.D., 0.10 μm (28343-U)
oven
80 °C (0.5 min), 30 °C/min to 280 °C (3 min)
inj. temp.
250 °C
detector
μ-ECD, 300 °C
carrier gas
hydrogen, 40 cm/sec, constant
injection
1 μl, splitless (0.75 min)
liner
4 mm I.D., single taper
sample
Ar1254 (44808) and surrogates (48460) at 500 ppb and 50 ppb in n-hexane