跳转至内容
Merck
CN

GC Analysis of Cocaine in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber

GC Analysis of Cocaine in Urine on a 100% Methyl Column after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

材料

used together

产品编号
说明
价格

SPME Fiber Assembly

100 μm PDMS, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, red hub

相关产品

产品编号
说明
价格

Equity-1 Capillary GC Column

L × I.D. 30 m × 0.32 mm, df 0.25 μm

CONDITIONS

sample/matrix

0.5 mL urine (250 ng each analyte, 20 μL 2.5% NaF) in 1 mL vial

SPME fiber

polydimethylsiloxane, 100 μm (57300-U)

extraction

immersion, 30 min

desorption process

3 min, 240 °C

column

100% methyl polysiloxane, 30 m × 0.32 mm I.D., 0.25 μm (Supelco equivalent, Equity-1, 28055-U)

oven

120 °C to 280 °C at 10 °C/min

inj. temp.

splitless (splitter opened after 1 min), 240 °C

detector

NPD, 280 °C

carrier gas

helium, 3 mL/min