Merck
CN

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity-1 after SPME using 85 μm Polyacrylate Fiber

EN 14110: GC Analysis of Methanol Impurity in Biodiesel on Equity™-1 after SPME using 85 μm Polyacrylate Fiber suitable for GC, application for SPME

材料

SPME 光纤

产品编号
说明
价格

聚丙烯酸酯 (PA) 固相微萃取萃取头

df 85 μm(PA, for use with manual holder, needle size 24 ga

分析色谱柱

产品编号
说明
价格

Equity-1 毛细管GC色谱柱

L × I.D. 30 m × 0.32 mm, df 1.00 μm

CONDITIONS

sample/matrix

2 g of B100 biodiesel containing 0.01% methanol and 0.0785% 2-propanol in a 4 mL vial

SPME fiber

PK3 SPME ASSY 85UM POLYACRYLATE FS 24GA MAN (WHITE) (57304)

extraction

headspace, 60 °C for 5 min

desorption process

250 °C for 0.75 min

column

Equity-1, 30 m x 0.32 mm I.D., 1.0 μm (28057-U)

oven

50 °C

inj. temp.

250 °C

detector

FID, 200 °C

carrier gas

helium, 30 cm/sec

liner

0.75 mm I.D., SPME type, straight design (unpacked)