GC Analysis of Amphetamines in Plasma on SLB®-5ms (20 m x 0.18 mm I.D., 0.36 μm) after SPME using 100 μm PDMS Fiber, Fast GC Analysis

GC Analysis of Amphetamines in Plasma on SLB®-5ms (20 m x 0.18 mm I.D., 0.36 μm) after SPME using 100 μm PDMS Fiber, Fast GC Analysis suitable for GC, application for SPME

材料

used together

产品编号
说明
价格

SPME Fiber Assembly

100 μm PDMS, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, red hub

CONDITIONS

SPME fiber

100 μm PDMS (57300-U)

extraction

headspace, 300 rpm stirring at 75 °C for 15 min

column

SLB-5ms, 20 m x 0.18 mm I.D. x 0.36 μm (28576-U)

oven

75 °C (1 min), 10 °C/min to 190°C, 25 °C/min to 300 °C (4 min)

MSD interface

300 °C

carrier gas

helium, 0.7 mL/min. constant

liner

0.75 mm I.D. SPME