GC Analysis of Phenols in Water on a 5% Phenyl Column after SPME using 85 μm Polyacrylate Fiber

材料
相关产品
产品编号
说明
价格
聚丙烯酸酯 (PA) 固相微萃取萃取头
85 μm Polyacrylate, Fused Silica (1 cm), needle size 24 ga, Manual Holder, pk of 3, white hubCONDITIONS
sample/matrix
50 ppb phenols in water, 27% NaCl, pH 2
SPME fiber
polyacrylate, 85 μm (57304)
extraction
20 min immersed in water with rapid stirring
desorption process
splitless, 280 °C (closed 3 min)
column
5% Phenyl Polysiloxane, 30 m × 0.25 mm I.D., 0.25 μm (Supelco equivalent, SLB-5ms, 28471-U)
oven
40 °C (4 min), 12 °C/min to 260 °C
detector
MS, Scan Range m/z = 45-465 at 0.6 sec/scan
carrier gas
helium, 40 cm/sec at 40 °C