HPLC Analysis of Albuterol Tablets – Assay on Chromolith® HighResolution RP-18 endcapped 150 x 4.6mm

CONDITIONS
sample preparation
120 μg/mL of Albuterol Sulfate RS prepared as follows. Transfer USP Albuterol Sulfate RS to a suitable volumetric flask, and add a volumne of Solution A corresponding to 60% of the flask volume. Sonicate for 5 minutes, and dilute with Methanol to volume.
column
Chromolith® HighResolution RP-18 endcapped, 150 x 4.6 mm (1.52023.0001)
mobile phase
[A] 10mL/L of glacial acetic acid in water; [B] 1.13g of sodium 1-Hexanesulfate in 1200mL water, add 12mL of glacial acetic acid; (40:60,:A:B)
flow rate
1.5 mL/min
pressure
2001 psi (138 bar)
column temp.
ambient
detector
276 nm cell: 11 μL
injection
25 μL
sample
Albuterol RS 30 μg/mL in diluent (30 μg/mL of Albuterol Sulfate RS in Diluent from Standard stock solution)