HPLC Analysis of Arbutin and Hydroquinone in Whitening Serum with Ascentis® Express C18

HPLC Analysis of Arbutin and Hydroquinone in Whitening Serum with Ascentis® Express C18

材料

used together

产品编号
说明
价格

suitable for HPLC

相关产品

产品编号
说明
价格

Ascentis ® Express C8,2.7μm HPLC 色谱柱

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

熊果苷

analytical standard

熊果苷

European Pharmacopoeia (EP) Reference Standard

对苯二酚

Pharmaceutical Secondary Standard; Certified Reference Material

熊果苷

phyproof® Reference Substance

Millex PVDF 针式过滤器

pore size 0.45 μm, diam. 13 mm, non-sterile, hydrophilic

CONDITIONS

column

Ascentis® Express C18, 2.7 µm 100 x 2.1 mm I.D. (53832-U)

mobile phase

[A] Water; [B] Methanol; A/B 10/90 (v/v)

gradient

isocratic

flow rate

100 µL/min

pressure

81 bar (1175 psi)

column temp.

25 °C

detector

UV, 220 nm (micro flow cell; 1.4 µL/7 mm)

injection

0.5 µL

sample/matrix

Standard solution (10 µg/mL): Weigh accurately 5 mg of each standard into a 50 mL volumetric flask and add 40 mL diluent. Sonicate for 5 min and fill up to mark with diluent. Pipette 5 mL of this solution to a 50 ml volumetric flask and fill up to volume with diluent.
Sample preparation: Place 1 g of homogenized serum in a 50 mL volumetric flask, add 30 mL of diluent and sonicate for 5 minutes. Fill up to volume with diluent. Transfer 3 mL of this suspension into a 25 mL volumetric flask and fill up to volume with diluent. Pass through 0.45 µm syringe filter before injection.

说明

一般描述

Bleaching or skin whitening serum is widely used to reduce melanin in the skin, and for such products analytical quality control is required. This report focuses on the testing of arbutin and the toxic by-product hydroquinone in formulated serum products, using LC-UV.
For the analysis of arbutin and hydroquinone in skin lightening cream, the goal of this work is to improve existing methods (e.g. Int. J. Appl. Sci. Eng., 2011.9.4) using shorter and smaller particle size columns instead of a 250 x 4.6mm RP-18e 5 µm column.

分析说明

Here an Ascentis® Express C18 100 x 2.1 mm, 2.7 mm is used. An LOD <5 µg/mL and LOQ <10 µg/mL for arbutin, and a LOD <0.5 µg/mL LOQ <1 µg/mL of hydroquinone was achieved using a UV detection.

其他说明

APP_226