Merck
CN

HPLC Analysis of Caffeine and Taurine on Ascentis® Express HILIC Column using Evaporative Light Scattering Detection (ELSD)

HPLC Analysis of Caffeine and Taurine on Ascentis® Express HILIC Column using Evaporative Light Scattering Detection (ELSD) application for HPLC

材料

分析色谱柱

产品编号
说明
价格

Ascentis® Express HILIC, 2.7 μm HPLC色谱柱

2.7 μm particle size, L × I.D. 5 cm × 3 mm

标准

产品编号
说明
价格

牛磺酸

≥99%

咖啡因

certified reference material, TraceCERT®, Manufactured by: Sigma-Aldrich Production GmbH, Switzerland

CONDITIONS

column

Ascentis Express HILlC, 5 cm x 3.0 mm I.D., 2.7 μm particles (53967-U)

mobile phase

[A] 100 mM ammonium formate, pH 3.0 with formic acid; [B] acetonitrile; (10:90, A:B)

flow rate

0.6 mL/min

column temp.

25 °C

detector

ELSD, 55 °C, 3.5 bar nitrogen

injection

2 μL

sample

1000 μg/mL in 10:90, water: acetonitrile

说明

分析说明

This application demonstrates the suitability of Ascentis Express HILIC for the efficient separation of caffeine and taurine. Elution is by aqueous normal phase (ANP) as an increase in organic will increase retention.

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany