HPLC Analysis of Glucosinolates on Chromolith® RP-18 endcapped

HPLC Analysis of Glucosinolates on Chromolith® RP-18 endcapped

材料

used together

产品编号
说明
价格

甲醇

gradient grade for liquid chromatography LiChrosolv® Reag. Ph Eur

乙腈

gradient grade for liquid chromatography LiChrosolv® Reag. Ph Eur

for chromatography (LC-MS Grade) LiChrosolv®

相关产品

产品编号
说明
价格

桂竹香苷 钾盐

phyproof® Reference Substance

Glucobrassicanapin potassium salt

phyproof® Reference Substance

芸苔葡糖硫苷 钾盐

phyproof® Reference Substance

芝麻菜苷 钾盐

phyproof® Reference Substance

屈曲花苷 钾盐

phyproof® Reference Substance

2(R)-羟基-2-苯乙基芥子油苷 钾盐

phyproof® Reference Substance

Gluconapin potassium salt

phyproof® Reference Substance

豆瓣菜苷钾盐 钾盐

phyproof® Reference Substance

金莲葡萄糖硫苷 钾盐

phyproof® Reference Substance

前告伊春 钾盐

phyproof® Reference Substance

白芥子硫苷 钾盐

phyproof® Reference Substance

黑芥子硫苷酸钾 钾盐

phyproof® Reference Substance

表前告伊春 钾盐

phyproof® Reference Substance

萝卜苷 钾盐

phyproof® Reference Substance

Chromolith® Column Coupler

column coupler

Millex亲水PTFE针式过滤器

pore size 0.45 μm, diam. 13 mm, non-sterile

Chromolith® HighResolution RP-18 封端

L × I.D. 100 mm × 2 mm HPLC column

CONDITIONS

column

Chromolith® HighResolution RP-18 endcapped 200-2 (2x 100-2 mm columns coupled with 1.51467)

mobile phase

[A] Water; [B] acetonitrile

gradient

At 0 min, 100%A at 95µL/min; in 6.7 min, to 95%A at 190 µL/min; in 10.0 min to 80%A at 363 µL/min; in 3.3 min to 20%A at 363 µL/min.

flow rate

95-363 µL (see gradient)

column temp.

Ambient

detector

UV, 229 nm (micro flow cell; 1.4 µL/7 mm)

injection

1 µL

pressure

55-190 bar (798-2755 psi) (A031)

说明

一般描述

Glucosinolates are plant secondary metabolites present in Brassicaceae family plants, including broccoli, cauliflower, horseradish, and cabbage. When plants are chewed, cut, or otherwise damaged, glucosinolates form mustard oils that give the plants their pungent flavor. These oils have a characteristic bitter flavor and can be used for crop protection or as an antibacterial in medicinal products.In this method for the determination of glucosinolates, a Chromolith® HighResolution RP-18 endcapped column was used with UV detection.