跳转至内容
Merck
CN

HPLC Analysis of Insulin Variants and Analogs on BIOshell A160 Peptide C18

HPLC Analysis of Insulin Variants and Analogs on BIOshell™ A160 Peptide C18 application for HPLC

CONDITIONS

column

BIOshell A160 Peptide C18, 15 cm x 2.1 mm, 2.7 μm particles (66905-U)

mobile phase

[A] 75:25 water (0.1% trifluoroacetic acid) : acetonitrile (0.1% trifluoroacetic acid); [B] 50:50 water (0.1% trifluoroacetic acid) : acetonitrile (0.1% trifluoroacetic acid); (67:33, A:B)

flow rate

0.5 mL/min

pressure

3600 psi (248 bar)

column temp.

75 °C

detector

UV, 215 nm

injection

0.5 μL

sample

mix of six insulin variants, 100 μg/mL in 0.1% trifluoroacetic acid

说明

分析说明

BIOshell A160 Peptide C18 is an ideal column for reversed-phase separations of small proteins to large peptide sequences. There is over 75% sequence similarity between all six of the insulin variants in the sample; therefore, separation of these variants is achieved through the subtle differences in the polarities of surface residues on the peptides.

法律信息

BIOshell is a trademark of Sigma-Aldrich Co. LLC