Zenix® SEC-100 Gel Filtration Column, 3μm

L × I.D. 30 cm × 4.6 mm, 3 μm particle size, stainless steel

CONDITIONS

column

Zenix SEC-100, 30 cm x 4.6 mm I.D., 3 μm particles (Z777006)

column

Zenix SEC-150, 30 cm x 4.6 mm I.D., 3 μm particles (Z777016)

column

Zenix SEC-300, 30 cm x 4.6 mm I.D., 3 μm particles (Z777028)

mobile phase

0.2 M sodium phosphate dibasic, pH 7.0 with phosphoric acid

flow rate

0.25 mL/min

pressure

1000 psi (69 bar)

column temp.

25 °C

detector

UV, 215 nm

injection

0.5 μL

sample

Waters BEH200 protein standard

说明

一般描述

Pore size is a kay parameter of resolution size exclusion chromatography. Matching the pore size to resolution of the sample components is critical in selecting a column. This application shows the impact of pore size on resolution of sample components as a function of molecular size

法律信息