HPLC Analysis of N-Linked Glycans on BIOshell Glycan

HPLC Analysis of N-Linked Glycans on BIOshell Glycan application for HPLC

材料

分析色谱柱

产品编号
说明
价格

BIOshell Glycan HPLC Columns

L × I.D. 10 cm × 2.1 mm, 2.7 μm particle size

CONDITIONS

column

BIOshell Glycan HPLC
Column (50993-U)

mobile phase

(A) Water with 100 mM ammonium formate, pH 4.5; (B) Acetonitrile

gradient

min, %B, Flow (mL/min): 0-74-0.4; 15-71-0.4; 25-55-0.4; 26-45-0.2; 30-45-0.2; 31-75-0.4; 35-75-0.4.

pressure

1130 psi (78 bar)

column temp.

50 °C

detector

Fluorescence (&#955ex = 260 nm, &#955em = 430 nm)

injection

3 μL

sample

in 70% mobile phase B

说明

分析说明

Separation pf a mix of 2-AB labled N-linked glycans released from human IgG was optimized on BIOshell Glycan HPLC Columns. Baseline resolutions for all of the individual glycans were observed in both mixes, except for the positional isomer pair G1F and G1F′ which were partially resolved. 