CONDITIONS
column
BIOshell™ Glycan HPLC
Column (50993-U)
mobile phase
(A) Water with 100 mM ammonium formate, pH 4.5; (B) Acetonitrile
gradient
min, %B, Flow (mL/min): 0-74-0.4; 15-71-0.4; 25-55-0.4; 26-45-0.2; 30-45-0.2; 31-75-0.4; 35-75-0.4.
pressure
1130 psi (78 bar)
column temp.
50 °C
detector
Fluorescence (λex = 260 nm, λem = 430 nm)
injection
3 μL
sample
in 70% mobile phase B
说明
分析说明
Separation pf a mix of 2-AB labled N-linked glycans released from human IgG was optimized on BIOshell™ Glycan HPLC Columns. Baseline resolutions for all of the individual glycans were observed in both mixes, except for the positional isomer pair G1F and G1F′ which were partially resolved.