HPLC Analysis of Proteins by Cation Exchange on Proteomix® WCX: Affect of Particle Size

HPLC Analysis of Proteins by Cation Exchange on Proteomix® WCX: Affect of Particle Size application for HPLC

材料

used together

产品编号
说明
价格

磷酸

≥85 wt. % in H2O, ≥99.999% trace metals basis

相关产品

产品编号
说明
价格

Proteomix® WCX-NP1.7 Weak Cation Exchange Column

L × I.D. 5 cm × 2.1 mm, 1.7 μm particle size, stainless steel

Proteomix® WCX-NP3 Weak Cation Exchange Column

L × I.D. 5 cm × 4.6 mm, 3 μm particle size, stainless steel

Proteomix® WCX-NP5 Weak Cation Exchange Column

L × I.D. 5 cm × 4.6 mm, 5 μm particle size, stainless steel

CONDITIONS

column

Proteomix WCX-NP5, 5 cm x 4.6 mm I.D., 5 μm particles (Z777193)

column

Proteomix WCX-NP3, 5 cm x 4.6 mm I.D., 5 μm particles (Z777185)

column

Proteomix WCX-NP1.7, 5 cm x 4.6 mm I.D., 5 μm particles (Z777177)

mobile phase

[A] 10 mM potassium phosphate, dibasic, pH 7 with phosphoric acid; [B] 0.3 M potassium phosphate, dibasic, pH 7 with phosphoric acid

gradient

0 to 100% B in 10 min, held at 100% B for 5 minutes

flow rate

0.5 mL/min

pressure

700 psi, 1200 psi, 3400 psi total pressure, respectively (48 bar, 83 bar, 234 bar total pressure, respectively)

column temp.

25 °C

detector

UV, 280 nm

injection

5 μL

sample

g/L ea: ribonuclease a, 0.38; cytochrome c, 0.15; lysozyme, 0.23; a-chymptrypsinogen, 0.23; in mobile phase A

说明

分析说明

Smaller particles can affect retention and peak widths in ion exchange chromatogaphy of proteins. Smaller particles provide higher total surface area, lower peak widths, and higher backpressure. With higher resolution afforded by the smaller particle size, greater chromatographic detail is evident with minor contaminants.

法律信息

Proteomix is a registered trademark of Sepax Technologies