Merck
CN

RP-HPLC Analysis of Substituted Phenethylamine Psychedelics on Ascentis® Express 2.7 μm RP-Amide

RP-HPLC Analysis of Substituted Phenethylamine Psychedelics on Ascentis® Express 2.7 μm RP-Amide application for HPLC

材料

分析色谱柱

产品编号
说明
价格

Ascentis® Express RP-Amide 2.7μm HPLC 色谱柱

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

标准

产品编号
说明
价格

25B-NBOMe hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

25I-NBOMe hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

25C-NBOMe hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

25I-NBF hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

25I-NBMD hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express RP-Amide 10cm x 2.1 mm I.D., 2.7 μm particles (53913-U)

mobile phase

[A] 20 mM Ammonium acetate pH 4.0 with Acetic  Acid; [B] ACN; (85:15, A:B)

gradient

15% B for 2 min; 15 to 30% B in 6 min, held at 30% B for 4 min

flow rate

0.6 mL/min

pressure

4757 psi (328 bar)

column temp.

35 °C

detector

UV, 280 nm

injection

5 μL

sample

25C-NBOMe HCl, 25B-NBOMe HCl, 25I-NBF HCl, 25INBMD HCl, 25I-NBOME HCl, 20 μg/mL in 90:10 mobile phase A:mobile phase B

说明

分析说明

This application demonstrates the suitability of Ascentis Express RP-Amide for the HPLC separation of substituted phenethylamine psychedelics (Bath Salts). This application shows the use of UV detection for this set of compounds.

法律信息