跳转至内容
Merck
CN

UHPLC Analysis of Flakka and Related Compounds on Ascentis® Express RP-Amide in Reversed-Phase Mode

UHPLC Analysis of Flakka and Related Compounds on Ascentis® Express RP-Amide in Reversed-Phase Mode application for HPLC

材料

相关产品

产品编号
说明
价格

3,4-Methylenedioxypyrovalerone HCl (MDPV) solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

MDPBP hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Naphyrone hydrochloride

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

α-Pyrrolidinovalerophenone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

α-Pyrrolidinopropiophenone hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

Ascentis® Express RP-Amide, 2 μm UHPLC Column

2 μm particle size, L × I.D. 10 cm × 2.1 mm

CONDITIONS

column

Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2 μm (51576-U)

mobile phase

[A] 0.1% formic acid in water; [B] 0.1% formic acid in acetonitrile

gradient

5% to 95% A in 6 min

flow rate

0.4 mL/min

pressure

6400 psi

column temp.

35 °C

detector

UV, 265 nm

injection

1 μL

sample

50 μg/mL in 0.1% formic acid in water:0.1% formic acid in acetonitrile (90:10)

说明

分析说明

The rapid gradient separation of the street drug Flakka (α-PVP) and related compounds is shown here on an Ascentis Express RP-Amide column in reversed-phase mode. Highest grade UHPLC solvents were used to supply low background interference and low particulate contamination for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany