Merck
CN

LC/MS Analysis of Ethanol Metabolites Ethyl Sulfate and Ethyl-β-D-Glucuronide on Ascentis® Express RP-Amide

LC/MS Analysis of Ethanol Metabolites Ethyl Sulfate and Ethyl-β-D-Glucuronide on Ascentis® Express RP-Amide application for LC-MS

材料

分析色谱柱

产品编号
说明
价格

Ascentis® Express RP-Amide 2.7μm HPLC 色谱柱

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

标准

产品编号
说明
价格

乙基-β-D-葡萄糖醛酸标准液

1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®

乙基硫酸钠标准液 钠盐

1.0 mg/mL in methanol (as ethyl sulfate), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express RP-Amide, 10 cm x 2.1 mm I.D., 2.7 μm particles (53913-U)

mobile phase

0.1% (v/v) formic acid in water

flow rate

0.4 mL/min

pressure

3066 psi (211 bar)

column temp.

35 °C

detector

MS, ESI(-), combined XIC, 125.0 m/z, 221.1 m/z

injection

2 μL

sample

2 μg/mL in 99:1, water:methanol

说明

分析说明

The compounds ethyl sulfate and ethyl-β-D-glucuronide are metabolites of ethanol. Their analysis is of clinical interest as biomarkers or indicators of alcohol-induced liver disease (ALD). The HPLC analysis of these compounds using reversed-phase mode using conventional C18 chemistries suffers from poor retention due to the compounds′ polar nature. However, polar compound retention is improved by using the Ascentis Express RP-Amide, a polar-embedded phase, while still keeping the reversed-phase mechanism. Cerilliant CRMs provided reliable quantification.

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany