instant oatmeal, ground
Supel-Select HLB SPE, 30 mg/1 ml (54181-U)
1 mL methanol, 1 mL 50:50 methanol:water
0.5 mL extract
collect the second loaded aliquot of the extract
filter into low absorption vials
apHera NH2, 15 cm x 4.6 mm I.D., 5 μm particles (56401AST)
[A] water; [B] 20 mM ammonium carbonate, pH 9; [C] methanol:water (50:50)
100% A for 2 min, to 90% B, 10% C in 0.1 min, held until 10 min, to 100% A in 0.1 min, held for 5 min
0.5 mL/min
3191 psi (220 bar)
35 °C
MS/MS, ESI negative, 168/63, 168/79
60 μL
Direct analysis of glyphosate with LC-MS detection using the presented method is possible for oat, corn and rice products. Polymeric SPE was used for sample cleanup. Polymer-based apHera NH2 HPLC column provided stable and robust separation at basic pH.
Sample pretreatment: 5 g of homogenized cereal was mixed wtih 10 mL water and let sit for at least 30 minutes. 10 mL 1% (v/v) formic acid in acetonitrile was added, shaken for 15 minutes on a laboratory shaker, and centrifuged.5 g of homogenized cereal was mixed wtih 10 mL water and let sit for at least 30 minutes. 10 mL 1% (v/v) formic acid in acetonitrile was added, shaken for 15 minutes on a laboratory shaker, and centrifuged.