LC/MS Analysis of Meglumine on SeQuant® ZIC-HILIC

CONDITIONS
column
SeQuant® ZIC-cHILIC, PEEK, 150 x 4.6 mm, 3 μm, 100Å (1.50661.0001)
mobile phase
[A] acetonitrile; [B] 50 mM ammonium acetate, adjusted with acetic acid to pH 5.0; (65:35, A:B)
flow rate
1.2 mL/min
pressure
1479 psi (102 Bar)
column temp.
45 °C (oven) (45 °C (detector cell))
detector
Refractive Index (RI) (cell: 10 μL)
injection
10 μL
sample
Dissolve 25 mg of meglumine standard in 10 mL of diluent (Diluent: Mobile phase)