LC/MS (TOF) Analysis of 2C Amines Drugs of Abuse in Urine on Ascentis® Express HILIC after Solid Phase Extraction (SPE) using Supel-Select SCX

LC/MS (TOF) Analysis of 2C Amines Drugs of Abuse in Urine on Ascentis® Express HILIC after Solid Phase Extraction (SPE) using Supel™-Select SCX application for LC-MS, application for SPE

材料

SPE 管或板

产品编号
说明
价格

Supel-Select SCX SPE 96-well Plate

bed wt. 30 mg/well, pk of 1

分析色谱柱

产品编号
说明
价格

Ascentis® Express HILIC, 2.7 μm HPLC色谱柱

2.7 μm particle size, L × I.D. 10 cm × 2.1 mm

标准

产品编号
说明
价格

4-Bromo-2,5-dimethoxyphenethylamine hydrochloride solution

1.0 mg/mL in methanol (as free base), ampule of 1 mL, certified reference material, Cerilliant®

2C-B-FLY hydrochloride solution

1 mg/mL in acetonitrile: water (9:1), ampule of 1 mL, certified reference material, Cerilliant®

2,5-Dimethoxy-4-(n)-propylthiophenethylamine hydrochloride solution

1.0 mg/mL in acetonitrile: water (9:1) (as free base), ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

sample/matrix

spiked urine, 1 mL

SPE well plate

Supel-Select SCX SPE 96-well Plate, 30 mg/well (575664-U)

condition

1 mL methanol, then 1 mL 0.1% formic acid in water

washing

2 x 1 mL 10% methanol in water, 2 x 1 mL 1% formic acid in methanol

elution

2 mL 5% ammonium hydroxide in acetonitrile

eluate post-treatment

evaporate to 500 μL with nitrogen at room temperature, reconstitute in 250 μL acetonitrile

column

Ascentis Express HILIC, 10 cm x 2.1 mm I.D., 2.7 μm particles (53939-U)

mobile phase

5 mM ammonium formate in acetonitrile:water (90:10, v/v)

flow rate

0.4 mL/min

pressure

1570 psi (108 bar)

column temp.

35 °C

detector

TOF/MS

injection

2 μL

说明

分析说明

The 2C amines are a family of illicit, psychedelic phenethylamines. The term "2C" refers to the two carbon atoms between the benzene ring and the amino group in the parent molecular structure. A selection of 2C amines is shown here extracted from urine using Supel-Select SCX polymeric SPE, separated on an Ascentis Express HILIC column, and detected by TOF-MS. Cerilliant CRMs were used to ensure reliable MS identification and quantification.