DART®-MS Analysis of Methamphetamine and Amphetamine in Urine Using BioSPME

材料
相关产品
产品编号
说明
价格
(±)-Amphetamine-d5 (deuterium label on ring) solution
100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®(+/-)-苯丙胺标准液 溶液
1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®(±)-甲基苯丙胺-D5标准液 CRM 溶液
100 μg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®(±)-甲基苯丙胺标准液 溶液
1.0 mg/mL in methanol, ampule of 1 mL, certified reference material, Cerilliant®CONDITIONS
SPME fiber
BioSPME (SPME LC) Mixed Mode fiber
extraction
10 min with 1 mL sample
desorption process
Direct MS
detector
MS, Full scan 100-500 amu
说明
分析说明
BioSPME was used to determine the presence of methamphetamine and amphetamine in urine samples. Initial screening results can be obtained in less than 60 seconds by analyzing the extracted urine samples on a DART-MS system. By employing a fast and accurate screening method, the number of samples that require chiral separation can be reduced and laboratories can increase throughput and decrease costs associated with their drug screening programs.
法律信息
DART is a registered trademark of JEOL USA, Inc.