Merck
CN

HPLC Analysis of B-Vitamins in Vitamin Water using Ascentis® Express RP-Amide and UV Detection

HPLC Analysis of B-Vitamins in Vitamin Water using Ascentis® Express RP-Amide and UV Detection application for UHPLC

材料

分析色谱柱

产品编号
说明
价格

Ascentis® Express RP-Amide 2.7μm HPLC 色谱柱

2.7 μm particle size, L × I.D. 10 cm × 3 mm

CONDITIONS

column

Ascentis Express RP-Amide, 10 cm x 3.0 mm I.D., 2.7 μm particles (53918-U)

mobile phase

[A] 20 mM potassium phosphate, pH 3; [B] methanol

gradient

held at 0.5% B for 1.5 minutes, to 25% B in 1.0 minutes, held at 25% B for 1.7 minute, held at 0.5% B for 1.2 minutes

flow rate

0.75 mL/min

pressure

5584 psi (385 bar)

column temp.

30 °C

detector

UV, 210 nm

injection

0.5 μL

sample

vitamin water was filtered through 0.2um nylon syringe filter

说明

分析说明

The vitamin water drink was filtered using 0.2 um Nylon filter and injected into UHPLC for B-vitamin analysis

法律信息

Ascentis is a registered trademark of Merck KGaA, Darmstadt, Germany