UHPLC Analyis of Protein Standards on Sepax Unix SEC-300: Column Performance as a Function of Flow Rate

UHPLC Analyis of Protein Standards on Sepax Unix™ SEC-300: Column Performance as a Function of Flow Rate application for UHPLC

材料

相关产品

产品编号
说明
价格

Unix SEC-300 UHPLC Column

L × I.D. 15 cm × 4.6 mm, 1.8 μm particle size

CONDITIONS

column

Sepax Unix SEC-300, 15 cm x 4.6 mm I.D., 1.8 μm particles (Z777300)

mobile phase

0.2 M sodium phosphate, dibasic, pH 7.0 with phosphoric acid

flow rate

0.1 mL/min

flow rate

0.3 mL/min

flow rate

0.5 mL/min

pressure

500 psi (34 bar)

pressure

1550 psi (107 bar)

pressure

2400 psi (165 bar)

column temp.

25 °C

detector

UV, 215 nm

injection

2 μL

sample

Waters BEH200 SEC protein standard

说明

分析说明

In considering the capabilities of modern analytical HPLC and UHPLC instrumentation and columns, column performance (as measured by peak efficiency) signifcantly declines as flow rate increases, in the case of size exclusion chromatography of proteins. Despite this universal phenomenon, due to the slow diffusion coefficients biomacromolecules, it is remarkable that resolution can remain more than adequate at the higher flow rates, and thus allow for dramatic gains in throughput, particularly with small-particle columns.