跳转至内容
Merck
CN
HomeChemical Analysis for Food and Beverage TestingAnalysis of Vitamins D2 and D3 in Infant Milk Powder

Analysis of Vitamins D2 and D3 in Infant Milk Powder by LC-MS/MS According to GB 5009.296-2023

Jack Wang
R&D APAC lab, Shanghai, China

Abstract

An isotope dilution LC-MS/MS method employing a UHPLC column packed with 2 µm superficially porous particles was developed for the simultaneous detection of vitamins D2 and D3 in infant milk powder according to the GB 5009.296-2023 standard. Saponification and solid-phase extraction were utilized to effectively overcome matrix challenges, allowing for a rapid and accurate quantification. The method fulfilled all requirements stated in GB 5009.296-2023 for linearity, reproducibility, recovery, and sensitivity.

Section Overview

Introduction

Two chemical structures drawn as skeletal formulas side by side on a white background. On the left is vitamin D2, also known as ergocalciferol, shown with a complex arrangement of fused rings and carbon chains. Its structure includes two six-membered rings and one five-membered ring, with double bonds and several methyl groups branching off. A side chain extends from the five-membered ring, containing a double bond and ending with a branched structure that has multiple methyl groups. Near the bottom of the molecule, a hydroxyl group is attached to a carbon chain extending from one of the rings. On the right is vitamin D3, known as cholecalciferol, which has a very similar core structure with the same three-ring system and a comparable carbon chain extending downward ending in a hydroxyl group. However, the side chain branching from the five-membered ring differs slightly: it lacks the double bond present in vitamin D2 and has a simpler chain with fewer methyl branches. Each structure is labeled below, with “Vitamin D2 (ergocalciferol)” under the left molecule and “Vitamin D3 (cholecalciferol)” under the right, clearly identifying the compounds.

Figure 1.Chemical structures of vitamin D2 (ergocalciferol, CAS: 50-14-6) & vitamin D3 (cholecalciferol, CAS: 67-97-0).

Vitamin D is a fat-soluble vitamin essential for regulating calcium metabolism and maintaining bone health. It exists in two forms: vitamin D2 (ergocalciferol) and D3 (cholecalciferol) (Figure 1). While both forms exhibit the same physiological effects, vitamin D3 is considered more effective at increasing the serum 25(H)D levels.1 Currently, the primary methods for vitamin D analysis include high-performance liquid chromatography and high-performance liquid chromatography-tandem mass spectrometry (LC-MS/MS).

In this study, isotope dilution LC-MS/MS using a 2 µm superficially porous particle UHPLC column was used to simultaneously detect vitamins D2 and D3 in infant milk powder. Infant milk powder typically contains low levels of vitamin D, and its complex sample matrix poses challenges for sample preparation and separation, making it potentially difficult to distinguish between vitamin D2 and D3.

Sample preparation was performed according to the GB 5009.296-2023 standard, “National Standards for Food Safety Determination of Vitamin D in Food”. Following saponification extraction, the infant milk powder extract was further cleaned up using silica gel-based solid-phase extraction (SPE) prior to analysis by LC-MS/MS. This method was found to effectively isolate trace amounts of vitamin D from complex matrices, thereby facilitating a rapid and accurate determination of vitamin D content in food samples such as the infant milk powder examined in this study.2-5

Experimental

The analysis was performed in accordance with the GB 5009.296-2023 method using the procedures detailed below.

Reagent Preparation

  • Potassium hydroxide solution (50 %): Dissolve 50 g potassium hydroxide in 50 g water, allow to cool to room temperature, and store in a polyethylene bottle.
  • BHT-ethanol solution (0.2 g/100 mL): Prepare freshly, by dissolving 1.0 g 2,6-di-tert-butyl-4-methylphenol (BHT) in 500 mL anhydrous ethanol.
  • Ethyl acetate - n-hexane solution (1+19): Mix ethyl acetate and n-hexane at a volume ratio of 1:19.
  • Ethyl acetate - n-hexane solution (3+17): Mix ethyl acetate and n-hexane at a volume ratio of 3:17.
  • Mobile phase A (0.05% formic acid in 5 mmol/L ammonium formate aqueous solution): Weigh 0.315 g ammonium formate, dissolve in 1000 mL water, add 0.5 mL formic acid, and mix well.
  • Mobile phase B (0.05% formic acid -5 mmol/L ammonium formate methanol solution): Weigh 0.315 g ammonium formate, dissolve in 1000 mL methanol, add 0.5 mL formic acid, and mix well.

Standard Preparation

Keep all prepared single-compound and mixed solutions at –18 °C, protected from light. Store according to the times indicated below.

Single component solutions

  • Vitamin D2 standard solution I (1000 mg/L): Accurately weigh 50 mg vitamin D2 standard into a beaker, dissolve in anhydrous ethanol, and transfer to a 50 mL volumetric flask. Adjust volume to the mark with anhydrous ethanol and mix well. Transfer the solution into an amber glass reagent bottle. Store for up to six months.
  • Vitamin D2 standard solution II (100 mg/L): Pipette 10.00 mL vitamin D2 standard solution I into a 100 mL volumetric flask, adjust volume to the mark with methanol, and mix well. Store for up to three months.
  • Vitamin D3 standard solution I (1000 mg/L): Accurately weigh 50 mg vitamin D2 standard into a beaker, dissolve in anhydrous ethanol, and transfer to a 50 mL volumetric flask. Adjust volume to the mark with anhydrous ethanol and mix well. Transfer the solution into an amber glass reagent bottle. Store for up to six months.
  • Vitamin D3 standard solution II (100 mg/L): Pipette 10.00 mL vitamin D3 standard solution I into a 100 mL volumetric flask, adjust volume to the mark with methanol, and mix well. Store for up to three months.
  • Vitamin D2-d3 standard solution (100 mg/L): Pipette 1.00 mL of a vitamin D2-d3 standard solution (1000 mg/L) into a 10 mL volumetric flask, adjust volume to the mark with methanol, and mix well. Store for up to three months.
  • Vitamin D3-d3 standard solution (100 mg/L): Pipette 1.00 mL of a vitamin D3-d3 standard solution (1000 mg/L) into a 10 mL volumetric flask, adjust volume to the mark with methanol, and mix well. Stored for up to three months.

Mix solutions

  • Mixed vitamin standard solution (vitamin D2 and D3 concentration 1.00 mg/L each): Pipette 1.00 mL vitamin D2 standard solution II and 1.00 mL vitamin D3 standard solution II into a 100 mL volumetric flask, adjust volume to the mark with methanol, and mix well. Store the solution for up to one month.
  • Mixed internal standard solution (Vitamin D2-d3 and vitamin D3-d3 concentrations 1.00 mg/L each): Pipette 100 μL each of vitamin D2-d3 standard solution and vitamin D3-d3 standard solution into a 10.0 mL volumetric flask. Fill to the mark with methanol and mix well. Stored the solution for up to one month.

Calibration solutions

Calibration solutions (nos. 1-5): Pipette 0.10 mL, 0.50 mL, 1.00 mL, 1.50 mL, and 2.00 mL, respectively, of mixed standard solution into separate 10 mL amber glass volumetric flasks. Add 1 mL of mixed internal standard solution to each flask. Adjust the volume to the mark with methanol and mix well. The final concentrations of vitamins D2 and D3 are 10.0 μg/L, 50.0 μg/L, 100 μg/L, 150 μg/L, and 200 μg/L, respectively, with each deuterated vitamin at 100 μg/L. Prepare these working solutions immediately prior to use.

Sample Preparation

  • Sample: Infant formula milk powder used for the study was obtained from a local supermarket and did not contain vitamins D2 and D3 (based on the label).
    Note: This sample was used as a blank sample for method validation experiments, and no background levels of vitamin D2 & D3 were detected in subsequent analysis.
  • Extraction: 1 g infant milk powder was weighed into a 50 mL centrifuge tube. Subsequently, 100 μL of the mixed internal standard solution, 0.4 g of ascorbic acid, and 10 mL water at 40-45 °C were added. The mixture was shaken for 1 min. A 10 mL BHT-ethanol solution (0.2 g/100 mL) was then added and shaken for 1 min, followed by the addition of 5 mL potassium hydroxide solution (50 %), and mixed thoroughly. Saponification was carried out in a thermostatic reactor at 80±2 °C for 30 min. Immediately after saponification, the mixture was cooled down to room temperature (25±5 °C) in a cold-water bath. To the cooled down mixture, 20 mL n-hexane was added, and the mixture was shaken for 3 min, followed by centrifugation at 8000 rpm for 5 min. The upper layer was transferred to a 50 mL centrifuge tube, 25 mL water was added, the mixture was shaken for 1 min, and then centrifuged at 8000 rpm. The final sample extract was collected as the upper hexane layer.
  • Cleanup: A Supelclean™ LC-Si SPE tube (6 mL/500 mg, 505374) was conditioned with 8 mL ethyl acetate, followed by 8 mL n-hexane. The sample extract was loaded onto the SPE tube and sequentially eluted with 6 mL of mixed ethyl acetate-n-hexane solution (1+19) and 6 mL of mixed ethyl acetate-n-hexane solution (3+17). The eluate was evaporated to dryness under nitrogen at 40 °C. The residue was reconstituted in 1.0 mL of methanol, ultrasonicated for 1 min, gently shaken for 1 min, and then filtered through a 0.22 μm PES filter membrane/syringe filter to obtain the final sample for analysis (Table 1).

Recovery and Precision

Two samples were prepared from separate 1.0 g portions of infant dairy product samples, which were spiked with 125 and 200 μL of the mixed vitamin standard solution, respectively. The resulting vitamin concentrations in these samples were 125 µg/kg and 200 µg/kg, respectively.

LC-MS/MS Method

The sample extract was analysed by LC-MS/MS using a 2 µm superficially porous particle Ascentis® Express 90 Å C18 column with multiple reaction monitoring (MRM).

Results & Discussion

Infant formula milk powder was extracted as described in the sample preparation section. Quantification of vitamins D2 (VD2) and D3 (VD3) was performed using an internal calibration (isotope dilution) approach with LC-MS/MS detection. The method performance was assessed by evaluating calibration linearity, precision, recovery, and sensitivity. Additionally, an actual infant formula milk powder sample was analyzed to demonstrate applicability.

The analysis on an Ascentis® Express 90 Å C18 2 µm column resulted in separation that was close to baseline under the selected conditions. The chromatogram obtained from the LC-MS/MS analysis of the vitamin D2 and vitamin D3 calibration solution 5 (200 μg/L) is shown in Figure 2, with the corresponding chromatographic data summarized in Table 2.

An LC-MS/MS total ion chromatogram (TIC) plotted with retention time in minutes on the x-axis, ranging from 0 to 15, and intensity in counts per second (cps) on the y-axis, scaling up to 1.0 times ten to the power of seven. A single chromatographic trace runs across the plot, nearly flat at baseline from 0 to about 9 minutes. At this point, two sharp and closely eluting peaks rise prominently. The first peak, labeled “1,” appears at 10.06 minutes and is slightly smaller, corresponding to vitamin D2 based on the quantitative MRM transition from mass 397 to 107. The second peak, labeled “2,” is slightly taller and elutes immediately after the first at 10.29 minutes, representing vitamin D3 with the transition from mass 385 to 107. After these peaks, the trace quickly drops back toward baseline and continues flat to the end of the run at 15 minutes. The graph line is drawn cleanly, clearly showing the narrow peaks that indicate well-resolved separation and detection of the two vitamins in the calibration solution at a concentration of 200 micrograms per liter. The axes are clearly labeled, and the fine detail of the two closely spaced peaks highlights the method’s precision.

Figure 2.LC-MS/MS TIC chromatogram obtained for vitamins D2 (peak 1) and D3 (peak 2) calibration solution 5 (200 μg/L).

Calibration

The results obtained for the injected external calibration solutions in the range of 10-200 µg/L are shown in Figure 3 and Table 3. The linearity (R2) was determined to be 0.9992 (Vitamin D2) and 0.9998 (Vitamin D3), meeting the GB 5009.296-2023 criteria of ≥ 0.9980.

A calibration curve graph plotting the response in counts on the y-axis, ranging from 0 to 8.0 times ten to the power of five, against the concentration of vitamin D2 in micrograms per liter on the x-axis, ranging from 0 to 200. Five data points are displayed as solid dots, each corresponding to the measured response at concentrations of approximately 10, 50, 100, 150, and 200 micrograms per liter. These points lie closely along a straight line drawn through them, indicating excellent linearity. Above the line is the equation of the calibration curve, shown as y equals 3695.5 times x plus 25264, along with an R-squared value of 0.9992, which demonstrates a highly linear relationship between concentration and detector response. The graph’s line increases consistently from left to right, reflecting a strong positive correlation. The axes are clearly labeled, with the x-axis titled “Concentration Vit D2 (µg/L)” and the y-axis titled “Response (counts),” and tick marks along both axes help illustrate the scale.
A calibration curve graph plotting the response in counts on the y-axis, ranging from 0 to 8.0 times ten to the power of five, against the concentration of vitamin D3 in micrograms per liter on the x-axis, which spans from 0 to 200. Five data points are displayed as solid dots, representing measured responses at concentrations of approximately 10, 50, 100, 150, and 200 micrograms per liter. These points align closely along a straight line that increases steadily from left to right, indicating a strong positive linear relationship. The line is accompanied by the regression equation, shown above it as y equals 3709.2 times x plus 19268, and an R-squared value of 0.9998, reflecting excellent linearity and precision of the calibration. The axes are clearly labeled, with the x-axis titled “Concentration Vit D3 (µg/L)” and the y-axis titled “Response (counts),” and both include evenly spaced tick marks.

Figure 3. Calibration curves for vitamins D2 & D3 obtained by the analysis of calibration solutions 1-5 (10, 50, 100, 150, and 200 μg/L).

Data Precision and Recovery

To determine method precision, a sample spiked at a concentration of 125 µg/kg for vitamin D2 and vitamin D3 was used (Figure 4). The precision results, summarized in Table 4, showed a relative standard deviation (RSD) of 2.63% for vitamin D2 and 4.41% for vitamin D3, meeting the GB 5009.296-2023 specification of ≤ 20%.

The results of the recovery experiment, conducted using a sample spiked at 200 µg/kg of vitamins D2 and D3 in infant formula milk powder, are presented in Table 5. Recoveries of 87.6% for vitamin D2 and 88.3% for vitamin D3 were obtained, complying with the GB 5009.296-2023 required range of 70–110%. The blank milk powder sample used in these experiments showed no detectable background of vitamins D2 & D3 (Figure 5).

An LC-MS/MS total ion chromatogram (TIC) plotted with retention time in minutes on the x-axis, ranging from 0 to 15, and intensity in counts per second (cps) on the y-axis, which scales up to 8.0 times ten to the power of six. The chromatogram trace begins near the baseline and rises into a sharp peak at around 2.5 minutes, then drops back down and continues with small fluctuations. Just after 9 minutes, two closely spaced peaks labeled “1” and “2” appear, with peak 1 slightly smaller and preceding peak 2, corresponding to the detection of vitamins D2 and D3. Following these, the trace shows several more pronounced peaks between 12 and 15 minutes, indicating the presence of additional compounds in the matrix. The line is consistently thin and smooth, showing clear separation of multiple components, with the labeled peaks distinctly rising above the background to mark the detection of the spiked vitamins.

Figure 4.LC-MS/MS TIC chromatogram obtained for an infant formula milk powder sample spiked at 125 µg/kg with vitamins D2 and D3.

Sensitivity

For the determination of sensitivity of the LC-MS/MS method, the baseline noise of a blank infant milk powder sample was used (Figure 5). A value of 3N/X was applied to calculate the limit of detection (LOD), and 10N/X was applied to calculate the limit of quantification (LOQ). The resulting LODs and LOQs for vitamins D2 and D3 achieved using this LC-MS/MS method met and exceeded the requirements specified in the GB method (Table 6).

A chromatogram graph displaying data from an LC-MS/MS TIC analysis. It features a white background with a green line tracing the intensity of detected compounds over time. The x-axis, labeled "Retention time (min)," spans from 0 to 15 minutes, while the y-axis, labeled "Intensity (cps)," ranges from 0 to 10,000,000 counts per second. The green trace shows multiple peaks, indicating the presence of various compounds. A prominent peak appears around the 12-minute mark, standing out as the highest point on the graph. Smaller peaks are scattered between 3 and 15 minutes, suggesting lower concentrations of other compounds.

Figure 5.LC-MS/MS TIC chromatogram obtained for an unspiked infant formula milk powder sample.

Conclusion

A method for the simultaneous determination of vitamins D2 and D3 in infant milk powder was established in accordance with the GB 5009.296-2023 standard. The sample was prepared using saponification, solvent extraction, and solid-phase extraction cleanup. Analysis was performed by UHPLC-MS/MS using an Ascentis® Express 90 Å C18 2 µm superficially porous particle column with stable isotope-labeled standards for internal calibration. The sample preparation method effectively eliminated interferences and reduced the matrix effects. The developed method satisfied all requirements specified in the GB 5009.296-2023 standard with respect to linearity, reproducibility, recovery, and sensitivity.

Related Products

LC Column & SPE

Loading

Solvents, Reagents & Consumables

Loading

Reference Materials

Loading

References

1.
Vitamin D. [Internet]. Nih.gov..[cited 04 Jun 2025]. Available from: https://ods.od.nih.gov/factsheets/VitaminD-HealthProfessional/
2.
Zhao K, Lou T, Chen Q, Ge B, Zheng W. 2015. Determination of Fat-soluble Vitamins in Infant Formula Milk Powder by Isotope Dilution/Liquid Chromatography-tandem Mass Spectrometry . Journal of Analytical Testing. 34(12):1372-1376. https://doi.org/10.3969/j.issn.1004-4957.2015.12.007
3.
Perales S, Delgado M, Alegría A, Barberá R, Farré R. 2005. Liquid chromatographic determination of Vitamin D3 in infant formulas and fortified milk. Analytica Chimica Acta. 543(1-2):58-63. https://doi.org/10.1016/j.aca.2005.04.019
4.
Zeng H, Zou X, Li Y, Xu J. 2008. Simultaneous determination of vitamin A,D3,E and β -carotene in health food by HPLC . J. Analysis Laboratory. 27(2):15-18. https://doi.org/10.3969/j.issn.1000-0720.2008.02.004
5.
National Standards for Food Safety Determination of vitamin D in food GB 5009.296-2023 . [Internet].
登录以继续。

如要继续阅读,请登录或创建帐户。

暂无帐户?