跳转至内容
Merck
CN

GC Analysis of Methylcyclohexane Methanol (MCHM) and Propylene Glycol Phenyl Ether (PPH) in Water on VOCOL® after SPME using 100 μm PDMS Fiber

GC Analysis of Methylcyclohexane Methanol (MCHM) and Propylene Glycol Phenyl Ether (PPH) in Water on VOCOL® after SPME using 100 μm PDMS Fiber suitable for GC, application for SPME

材料

used together

产品编号
说明
价格

SPME Fiber Assembly

100 μm PDMS, Fused Silica (1 cm), needle size 24 ga, Autosampler, pk of 3, red hub

相关产品

产品编号
说明
价格

1-Phenoxy-2-propanol

≥93%

VOCOL® Capillary GC Column

L × I.D. 20 m × 0.18 mm, df 1.00 μm

配件

产品编号
说明
价格

Inlet Liner, Direct (SPME) Type, Straight Design (unpacked)

pkg of 1 ea

Headspace vial, screw top, rounded bottom (vial only)

volume 10 mL, clear glass vial, thread for 18, O.D. × H 22.5 mm × 46 mm, pkg of 100 ea

Magnetic Screw Cap for Headspace Vials, 18 mm thread

PTFE/silicone septum (white PTFE/tranparent blue silicone), septum thickness 1.3 mm, pkg of 100 ea

CONDITIONS

sample/matrix

MCHM and PPH at 1 μg/L in 6 mL water, 25% sodium chloride

SPME fiber

SPME Assembly, 100 μm PDMS FS, 24 ga, Auto (Red) (57301)

extraction

5 min, headspace at 55 °C, with agitation

desorption process

2 min, 270 °C (splitter open at 0.75 min)

column

Vocol, 20 m x 0.18 mm I.D. x 1.0 μm (28463-U)

oven

50 °C (1 min), 15 °C/min to 220 °C (5 min)

inj. temp.

270 °C

detector

MSD, selected ion mode

MSD interface

220 °C

scan range

selected ion mode; m/z=110, 97 (MCHM) and m/z=94, 152 (PPH)

carrier gas

helium, 1 mL/min constant

injection

SPME

liner

0.75 mm I.D., SPME

说明

分析说明

In this application, an SPME method was developed for the analysis of two compounds involved in the chemical spill into the Elk River in West Virginia, which took place in January 2014. The method could easily detect both MCHM and PPH at a level of 1 μg/L from spiked surface water.

法律信息

VOCOL is a registered trademark of Merck KGaA, Darmstadt, Germany