LC/MS/MS Analysis of Vitamin D Metabolites on Ascentis® Express F5 (Monitored Ions)

LC/MS/MS Analysis of Vitamin D Metabolites on Ascentis® Express F5 (Monitored Ions) application for HPLC

材料

used together

产品编号
说明
价格

Ammonium formate

eluent additive for LC-MS, LiChropur, ≥99.0%

相关产品

产品编号
说明
价格

1α-Hydroxyvitamin D3

≥97.0% (HPLC)

3-epi-25-Hydroxyvitamin D3

98% (CP)

Ascentis® Express F5, 2.7 μm HPLC Column

2.7 μm particle size, L × I.D. 15 cm × 2.1 mm

3-epi-25-Hydroxyvitamin D2 solution

100 μg/mL in ethanol, 98% (CP)

25-Hydroxyvitamin D2 solution

50 μg/mL in ethanol, ampule of 1 mL, certified reference material, Cerilliant®

25-Hydroxyvitamin D3 solution

100 μg/mL in ethanol, ampule of 1 mL, certified reference material, Cerilliant®

1α,25-Dihydroxyvitamin D3 solution

5 μg/mL in ethanol, ampule of 1 mL, certified reference material, Cerilliant®

1α,25-Dihydroxyvitamin D2 solution

5 μg/mL in ethanol, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

column

Ascentis Express  F5, 15 cm x 2.1 mm I.D., 2.7 μm particles (53571-U)

mobile phase

[A] 5 mM ammonium formate water; [B] methanol; (25:75, A:B)

flow rate

0.4 mL/min

pressure

5178 psi (357 bar)

column temp.

40 °C

detector

ESI(+), 100-1000 m/z

injection

2 μL, each compound 300 ng/mL in 40:60 (water:methanol)

说明

分析说明

The metabolic role and health implications of the various vitamin D isoforms are of current clinical interest. Therefore, it is important to have an analytical method that will resolve all of the known isoforms with necessary sensitivity and specificity.This application shows that a combination of Ascentis Express F5 with MS detection provided rapid resolution of seven such compounds. The highest grade LC-MS solvents were used to supply low background interference and low particulate contaminants for robust, trouble-free operation. Cerilliant CRMs provided reliable identification and quantification.