LC/MS with Online Solid Phase Extraction (SPE-LC/MS) of Thyroid Hormones in Plasma on Ascentis® Express Phenyl-Hexyl using RP-Amide Traps

LC/MS with Online Solid Phase Extraction (SPE-LC/MS) of Thyroid Hormones in Plasma on Ascentis® Express Phenyl-Hexyl using RP-Amide Traps application for LC-MS, application for SPE

材料

used together

产品编号
说明
价格

Ascentis® Express保护柱套

for use with Ascentis Express Guard Columns, pk of 1

Ascentis® Express RP-Amide,2.7 μm保护柱芯

2.7 μm particle size, L × I.D. 5 mm × 2.1 mm, pkg of 3 ea

相关产品

产品编号
说明
价格

OPTI-SOLV EXP 在线过滤器

holder with EXP titanium hybrid ferrule, pack of 1 sets

EXP® Pre-Column Filter Cartridges

0.2μm, pkg of 5 ea

Ascentis® Express Phenyl-Hexyl,2.7μm HPLC色谱柱

L × I.D. 5 cm × 2.1 mm, pk of 1 ea

L-甲状腺素(T4) 溶液

100 μg/mL in methanol with 0.1N NH3, ampule of 1 mL, certified reference material, Cerilliant®

3,3′,5-三碘-L-甲状腺原氨酸(T3) 溶液

100 μg/mL in methanol with 0.1N NH3, ampule of 1 mL, certified reference material, Cerilliant®

3,3′,5′-三碘代-甲腺原氨酸(rT3)标准液 溶液

100 μg/mL in methanol with 0.1N NH3, ampule of 1 mL, certified reference material, Cerilliant®

CONDITIONS

sample preparation

SPE (Solid Phase Extraction)

column

Ascentis Express Phenyl-Hexyl, 5 cm x 2.1 mm I.D., 2.7 μm particles (53334-U)

mobile phase

[A] water; [B] methanol, both with 0.1% acetic acid

gradient

5% B for 3 min; to 70% B in 2 min, hold 3 min

pressure

1800 psi

column temp.

ambient

detector

MS, ESI(+), MRM mode (777.7/731.8 for T4; 651.8/605.5 for T3 or rT3)

injection

5 μL

sample

protein precipitated rabbit plasma spiked with desired amount of T4, T3 and rT3

说明

分析说明

An online SPE-LC/MS method is shown here for the determination of thyroid hormones in blood plasma. The method exploits RP-Amide or C8 as the trapping column and a phenyl-hexyl phase as the analytical column, respectively. Under the optimized conditions, both RP-Amide and C8 effectively trapped the thyroid hormones extracted from spiked rabbit plasma sample which had been protein precipitated. However, the RP-Amide traps were shown to be superior to the commonly used C8 traps because they provided higher analyte signals and recoveries. Additionally, RP-Amide traps are more flexible with washing solvent as they are compatible with 100% aqueous mobile phases and can be used with up to 20% methanol with minimal sample loss. Separation was carried out using an Ascentis Express 2.7 μm Phenyl-Hexyl column. The highest grade solvents provided clean, robust operation. Cerilliant CRMs provided reliable quantification.